MODEL MFS-01

FOOT-SHOCK STRESS SYSTEM


MFScage_rm.jpg MFS_con.jpg

Introduction

MFS-01 was designed to give a stress to the animal in a cage by giving it an electrical shock through the grid floor. The system consists of a RSK-001 rat cage (or a MSK-001 mouse cage), a SGS-003 shock generator scrambler, a CBX-CT cycle timer, a CSC-001 CS Controller and a MCS-01 Clip-Mount CS Presentation Unit.


SPECIFICATIONS

 RSK-001 RAT CAGE
RSK001.jpg  Grid  5mm O.D. x 18 , 16mm Interval (Center-Center) 
 Material  Stainless Steel, Acryl
Dimensions  OD 360W x 370D x 450mmH
 ID 300W x 300D x 400mmH
Weight  Approx. 3kg


 MSK-001 MOUSE CAGE
MSK001.jpg  Grid 3mm O.D. x 24, 10mm Interval (Center-Center)
 Material Stainless Steel, Acryl
Dimensions OD  300W x 275D x 290 mmH 
ID  200W x 240D x 230 mmH 
Weight Approx.  2kg


SGS-003 SHOCK GENERATOR SCRAMBLER
 Shock Current  0.5, 0.6, 0.8, 1.0, 1.3, 1.6, 2.0, 2.5, 3.0, 4.0mA RMS
 Range  x1, x0.1 
 Compliance Voltage  AC 1400Vp-p  50/60Hz
 Scramble  9 line multi-scramble by logic control
 Input  5-24V DC or Contact
 Output Timer  Remote Mode : output time control by external input
 Output Connector  20P square connector for scramble output
 Dimensions  300W x 270D x 100 mmH 
 Weight  Approx. 5kg
 Power  100V AC 50/60Hz

sgs003n.jpg
SGS-003 Shock Generator Scrambler

Newly developed SGS-003 Shock Generator Scrambler has made it possible to give an animal a stable electrical shock by the adoption of the high-voltage AC type and logically controlled multi-scrambler. The shock current will be precisely delivered to an animal on the floor grid even if it is a short-time stimulation.

Output current can be set for the range of 50 uA-4mA and output duration can be controlled by CBX-CT Cycle Timer.



CBX-CT CYCLE TIMER
 Setting  By digital SW, toggle SW
 Digital SW  TIMES  1-1000  Number of Cycle
 ITI  1-199  Waiting Period (sec)
 CS1  0-190  Warning Period (sec)
 CS2  0-19  Warning Period (sec)
 UCS  0.1-19  Stimulation Period (sec)
 Toggle SW

  TONE(CS1, CS2, UCS) 
   ON/OFF of TONE output 

  LIGHT(CS1, CS2, UCS) 
   ON/OFF of LIGHT output 

 START SW  Starts the preset rogram
 STOP SW  Stops the program (Double Click)
 Display  LED ( RUN, TONE, LIGHT, SHOCK )
Output  Open Collecter (30V DC 30mA)
 Dimensions  320W x 230D x 120 mmH
 Weight  Approx. 3kg
 Power  100V AC  50/60Hz  1A

CBX-CT.jpg
CBX-CT CYCLE TIMER

Random cycle function is included.



CSC-001 CS CONTROLLER
CSC001.jpg CS1  Stimulus Tone(2.9Hz Tone Generator)
 Tone volume adjustable with variable resistor (24VDC) 
CS2  Stimulus Light (AC100V)
 Electric power adjustable with variable resistor (200W max.) 
Dimensions  320W x 230D x 80mmH 
Weight  2.5 kg
Power  100VAC 50/60Hz 3A


  MCS-01 CLIP-MOUNT CS PRESENTATION UNIT  
MCS01.jpg  Height   60 cm 
 Weight   1 kg 



Specifications are subject to change without notice.


Last Updated : November 20, 2007